Ti-40 Ti-50 Potentiometric Titrator
Description
Ti-40
Potentiometric titrator with LCD display and smart guide system
Interchangeable burette: 10ml or 20ml selectable
Titration modes: DET, MET, SET and MAT
pH calibration and measurement supported
Stores up to 50 titration data sets (GLP-compliant)
RS-232/USB communication for PC control and printer output
Burette resolution: 1/14000, Repeatability: 0.2%
1. 【LCD Display & Smart Guide System】 Features LCD display screen with an intelligent guide system that provides step-by-step operation guidance, making the titrator easy to use for both experienced and novice operators.
2. 【Interchangeable High-Accuracy Burette】 Supports replaceable burette with two volume options: 10ml (accuracy ±0.025ml) and 20ml (accuracy ±0.035ml), providing flexibility for different titration applications. Burette resolution reaches 1/14000 with repeatability of 0.2%.
3. 【Multiple Titration Modes】 Supports four titration modes to meet diverse analytical requirements: DET (dynamic equivalence point titration), MET (monotone equivalence point titration), SET (preset endpoint titration) and MAT (manual titration), offering maximum flexibility for different sample types and methods.
4. 【pH Calibration & Measurement】 Built-in pH calibration and measurement function supports both pH and mV measurements, expanding application range beyond traditional titration analysis.
5. 【GLP-Compliant Data Management】 Stores up to 50 titration data sets in compliance with GLP requirements, along with the latest titration curve. Ensures data integrity, traceability and regulatory compliance for quality control laboratories.
6. 【High-Precision Measuring Unit】 Measuring unit range covers (-1800.0~1800.0) mV and (0.00~14.00) pH, with resolutions of 0.1mV and 0.01pH. Delivers pH accuracy of ±0.01pH and mV accuracy of ±0.05%FS, with fluctuation of only ±0.3mV/3h for exceptional stability.
7. 【Automatic Temperature Compensation】 Built-in temperature compensation function with range of -5.0~105.0°C, resolution of 0.1°C and accuracy of ±0.3°C, ensuring accurate results across varying temperature conditions.
8. 【Flexible Data Connectivity】 Equipped with RS-232 communication interface for easy data transfer to printers, and USB/RS-232 interfaces for computer control, enabling comprehensive data management and automated operation.
9. 【Robust & Compact Design】 With dimensions of 340×400×400mm and weight of 10kg, the instrument features a stable benchtop design suitable for demanding laboratory environments.
10. 【Wide Applications】 Ideal for potentiometric titration in pharmaceutical, chemical, food, environmental and educational laboratories. Suitable for acid-base titration, redox titration, precipitation titration and complexometric titration applications.

Specifications
| Ti40 | Ti50 | |
| Measure range |
pH: (0.00~14.00)pH mV: (-1800.0~1800.0)mV Temp.: (-5.0~105.0)℃ |
pH: (0.00~20.00)pH mV: (-2000.0~2000.0)mV Temp.: (-5.0~105.0)℃ |
| Resolution |
pH value: 0.01pH mV value: 0.1mV Temperature: 0.1℃ |
pH value: 0.001pH mV value: 0.01mV Temperature: 0.1℃ |
| data storage | 50 sets | 200 sets titration results and a set of titration curve |
| Interface | RS-232 | USB interface, RS-232 printer port, special titration software |
| Applied range | widely used for the pharmaceutical industry, food industry, petrochemical industry, metallurgy, environmental protection, electroplating materials industry | Widely used in food additives, food safety, pharmaceutical production, petrochemical, quality control, research institutions |
| Advantages |
1. With pre-titration, endpoint titration preset, blank titration or manual titration functions and can be used to generate special titration according to user mode 2. Different electrode can be used: acid-base titration, redox titration, precipitation titration, complexometric titration, and other non-aqueous titration and pH measurement 3. Mixing system using PWM modulation technology, low noise |
1. The compact designed integrated titrator and convenient automatic titration system 2. GLP compliant measurement information, can be traced back more complete information 3. Upgrade function, reduce costs titration, make the operation more perfect 4. Titrator control: dot matrix LCD. Key operation. While controlling software. Or are controlled |